Looks Series (Surgical Telescope) www.xenosys.co.kr
9
Finally, please tighten the screw by turning clockwise at the screw (-) groove (part A) by
using the (-) driver.
④ To reset the PD, after turning the screw (-) groove to about 5~10°counterclockwise by
using the (-) driver, please tighten the knob(part B) by turning clockwise while holding the
driver at the screw groove(part A).
Caution: Do not overtighten the screw and knob as will wear out.
4) Adjustment of the pupil height
① Please place the Flip-Up telescope after the oculars should be in contact with eyeglass
lens or closely.
② Please check the clarity of vision by the object located at the WD (working distance).
③ Please adjust in the manner that the nose pad can be narrowed and/or temple arm can be
descended downward if the upper end part of the observed visual field through the
telescope is unclear, or spread the nose pad and/or fit the temple arm upward if the
lower end part is unclear.
Downward adjustment
Upward adjustment
5. Operating Instruction
A. After adjustment, take and use the surgical telescope. Some professionals with telescope
experience should use the telescope full-time, but most of first users could have experience
of such disorientation normally and require a break-in period. Depending on the
magnification level of telescope or personality, the break-in period should differ and does not
last more than two to three weeks normally. You had better to increase gradually the time of
daily use.
B. If your telescope is equipped with side shields, place the shield over temple. (Optional)
Not provided in Flip-Up type.
Installation to PT
Frame
Installation to HT Frame
Insert → Completed
Fit side shield in temple
→ Fix by Stopper →
Completed
Push
Push
Push
Completed