BRIEF INSTRUCTIONS 

CLEANING AND DISINFECTION OF 

DELTASTREAM HC

Xenios_HC_EN_003_2018_02 

Company 

Product name 

Actve substance 

Isopropanol

Isopropanol 

(70%)

Bode Chemie 

GmbH 

KohrsolinTM 

SteriliumTM 
Bacilol PlusTM 

Aldehyde 

Propanol 
Propanol 

Ecolab 

Healthcare 

GmbH

Incidin SchaumTM 
Incidin Spray 
LiquidTM 

Propanol/Ethanol 
Propanol/Ethanol 

B.Braun Mel-

sungen AG

MeliseptolTM 

Propanol 

Clorox Com-

pany USA 

Clorox Healthcare 

Bleach Germicidal 
WipesTM 

Hypochlorid 

Schülke&Mayr 

Mikrozid Liquid AF 

Propanol/Ethanol 

Metrex

Cavi 

Wipes

Isopropanol/Butox

yethanol 

PDI 

Super Sani Cloth 

Isopropanol 

We recommend changing the water with the subsequent 

disinfection of the water cycle as follows in order to kill 

water-borne pathogens (including non-tuberculous my-

cobacteria) and other germs:

 

• After each therapy,

• As part of the initial commissioning, 

• at latest 30 days after the start of therapy,

in case of storage and / or prolonged non-use 

Generally after 30 days, also with stored devices 

• every 12 months as part of the safety inspection.

• Replace hose set and couplings every 12 months.

Changing water and disinfection the water 
circuit

Filling the device during initial operation and recommis-

sioning: 

Fill approximately 1 liter of sterile water, e.g. with 

the help of a disinfected funnel into the water tank 

of the device.

 

Make sure that no foreign objects enter the water 

tank.

 

Observe the water level indicator during filling. 

The water level should be slightly below the MAX 

mark after filling.

 

After filling, screw the screw cap of the water filler 

neck back in until the screw cap seals. 

Check the microbial contamination regularly. We 

recommend to test the water quality once a 

month, nevertheless the system is in use or stored. 

Emptying the device: 

Disconnect the device from the mains.

 

Place a container (bucket, bowl, etc.) under the de-

vice or place it on a drain (e.g. sink).

 

Unscrew the cap of the water filler neck in sterile con-

dition.

 

Drain the tank contents into the collector via the con-

nected hoses. To do this, remove the adapter be-

tween the Hansen couplings. Make sure that the 

hoses are also completely emptied afterwards.

 

Table1: List of approved surface disinfectants*

*Maybe some of the listed products are not available

 

i

n your country. Take care about the availability!

 