Xavant STIMPOD NMS450 Quick Start Manual Download Page 2

When the patient is sedated and 

un

relaxed

1

5

6

7

2

3

4

Turn the device on by 
pressing the power button

Adjust the current (if needed) 
with the adjustment wheel. 
Press the ENTER button to 
confirm the new current.

Connect the ECG clips to the 
NMBA electrode. Place the 3D 
accelerometer on the patient’s 
thumb, using the strap to 
secure it. Make sure there is 
enough space for the thumb to 
move freely.

Press the play button to 
initiate the TOF sequence.

The actual current delivered (1) and the 
TOF Ratio (3) will be displayed on the 
screen. A refractory timer (2) will indicate 
the time needed for the nerve to recover 
before another stimulation is allowed.

Place the NMBA electrode on 
the patient’s ulnar nerve. The 
placement of the electrode can also 
be confirmed by using the mapping 
function on the STIMPOD.

Connect the NMBA cable to the 
STIMPOD.

Recall the last current setting by 
pressing the enter button.

This is what will be displayed 

This is what will be displayed 