Adjusting Settings

Adjusting the Current 

Current Range: 

0 – 30 mA adjustable in 1mA increments

Circle the Wheel to adjust the current

 

1.7)     Countdown Timer

  

Start the stimulation by pressing the play/pause button. 

 

The device will automatically start stimulation at the set    

  

  

frequency and the countdown timer will start.

 

The countdown will start from the time set in the 

 

    “Countdown Timer” option in the main menu. 

 Pause the timer and the stimulation by pressing the 

 

    play/pause button. 

 Reset the timer by pressing and holding the play/pause  

    button for longer than 2 seconds. 

   The Countdown Timer defaults at 10 minutes. This setting can  

    be changed in the Setup Menu.

 The Stimulation will stop when the timer reaches 00:00.  

    An alarm will sound and the Countdown Timer will reset.

Adjusting the Pulse Width

Options in the Setup Menu: 

0.1ms, 0.2ms

Default: 

0.1ms

Press 

Menu/Pulse

 Width button to toggle between  

different Pulse Widths.

Adjusting the Stimulation Frequency

Options in the Setup Menu: 

1Hz, 2Hz, 5Hz, 10 Hz

Default: 

2Hz

Press 

Enter/Hz

 button to toggle between different stimulating frequencies.

1.6) 

Open Circuit Detection

A few milliseconds prior to the actual stimulation, an impedance measurement 
is made to detect whether the connection between the STIMPOD NMS460 and 
the patient comprises a closed circuit.

Closed Circuit Detected:

 

Stimulation

 will take place.

 

Stimulating sound

 will be heard 

 The 

LED

 stimulus indicator will pulse 

green

 with every successful stimulus atempt.

 The 

information screen 

will show the value of the actual current delivered.

Open Circuit Detected:

 

No stimulation

 will take place.

 

No stimulating sound

 will be heard.

 The 

LED

 stimulus indicator will pulse 

red

 with every  

unsuccessful stimulus attempt.

 A 

warning screen

 will appear in the diagnostics screen indicating that an open 

circuit was detected.

1.8)   Shutdown

Auto Shutdown

 

STIMPOD NMS460 will automatically switch off after 10 minutes of no user or 

patient interaction.

Battery Depleted

 

When the battery is depleted the unit will prompt the user to replace the  
battery and switch off.

1.9)   Symbols 

Caution

Type BF
Applied Part

Separate collection 
for electrical and 
electronic equipment
(Applicable to EU 
community only)

Manufacturing 

Date (Year)

Prescription 
use only

Serial Number

Non-sterile

Temperature 
limit

50°C

0°C

90%

10%

Humidity 
limitation

Manufacturer

Representative 
in the EU

Catalogue 
Number

Consult instructions 
for use

4

Summary of Contents for STIMPOD MMS460

Page 1: ...PRODUCT CODE XT 46006 EN XM400 31A04 07...

Page 2: ...i...

Page 3: ...be observed to verify normal operation in the configuration in which it will be used The patient should avoid contact with metallic objects that are grounded produce an electrical conductive connectio...

Page 4: ...e stimulators are a symptomatic treatment and as such suppresses the sensation of pain that would otherwise serve as a protective mechanism Effectiveness is highly dependent upon patient selection by...

Page 5: ...healthcare pro fessional only This equipment system may cause radio interference or disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or re...

Page 6: ...e d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separatio...

Page 7: ...pplicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separa...

Page 8: ...er 4 1 8 Shutdown 4 1 9 Symbols on device battery clip 4 2 Setting up Device Defaults 5 2 1 Languages 5 2 2 Stimulation Mode 5 2 3 Pulse Width Options 6 2 4 Stimulating Frequency 6 2 5 Countdown Timer...

Page 9: ...n those supplied with the STIMPOD NMS460 may result in serious injury NOTE Standard ECG Electrodes can be used for the anode Such ECG electrodes need to conform to ANSI AAMI EC12 on disposable ECG ele...

Page 10: ...s to Enter in setup menu Stimulating LED indicator Flashing Green Stimulus delivered Flashing Red Open Circuit detected Menu Pulse Width Button Press to toggle between Pulse Widths Press and hold to a...

Page 11: ...too high higher than 13 kOhm This is most likely caused by a poor grounding electrode skin connection Pause This warning informs the user that the PAUSE button was pressed The STIMPOD NMS460 will pau...

Page 12: ...ircuit Detection A few milliseconds prior to the actual stimulation an impedance measurement is made to detect whether the connection between the STIMPOD NMS460 and the patient comprises a closed circ...

Page 13: ...croll down the menu to EXIT MENU and press enter A conformation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you to the previous...

Page 14: ...ter A confirmation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you back to the previous screen 2 5 Countdown Timer DEFAULT 10 mi...

Page 15: ...rstheusertheoptiontoenteruserinformation Twolinesof20 characters each may be entered This user information will be displayed for two seconds when unit is switched on as per example Use the Wheel to na...

Page 16: ...ld flash GREEN and if sound is enabled in the menu a beep should be heard each time a stimulus is delivered Stimulus should occur at the set frequency 1 2 or 5 Hz Use the adjusting wheel and slowly in...

Page 17: ...er Consumption 17mA Weight 130g Dimensions 145mm x 90mm x 30mm Operating Temperature 10 40 Celsius Storage and Transport Temperature 0 50 Celsius Operating Humidity 90 Relative Humidity 3 3 Cleaning a...

Page 18: ...e NMS 460 PERIPHERAL NERVE STIMULATOR ONLY XT 46000 No cables or accessories NMS 460 PERIPHERAL NERVE STIMULATOR KIT XT 46001 Including Mapping Probe Carry Case IFU NERVE MAPPING CABLE XT 46003 CARRY...

Page 19: ...11...

Page 20: ...Unit 102 The Tannery Industrial Park 309 Derdepoort Rd Silverton Pretoria South Africa 0184 Tel 27 0 12 743 5959 Fax 27 0 86 547 0026 E mail support xavant com Web www xavant com...

Reviews: