Instruction Manual
FH 10
– 500 EXT
8. CE Declaration of conformity
FH Adjustment Procedure
FH_10-500-BA-e-1213
6
1 Start the
instrument
Press ON/OFF
The green
light is on
2 Enter the
adjustment
procedure
Directly after pressing
ON/OFF, press PEAK
and PRINT together for
more times and
very
short intervals
until the
red light is on.
The lower
red light is
on
3 Instrument
Type
Press SET
Directly after pressing
PEAK + PRINT
XXX N is
shown
a
(Back to the
normal
mode ???)
(If you are in the
operation mode again,
the instrument has to be
shut off. Start again with
1. Eventually pressing
the buttons faster)
4 Select the
instrument
type
Press
▼▲
The
correspondi
ng value to
the instr.
appears in
the display
5 Save
instrument
type
Press SET
6 Enter the
adjustment
sequence
Press MEMORY
The upper
red light is
on
7 Chose the
adjustment
weight
Press UNIT and
with
▼▲ specify the
calibration weight in
Newton (X kg * 9,81)
The weight
is written in
the display
8 Save the
adjustment
weight
Press SET and UNIT
simultaneously
9 Attach the
weight to the
instrument
(by hanging)
Press ZERO
The
instrument
changes to
operation
mode.
Adjustment
procedure is
finished