C i 7 X X 0 B E N C H T O P S P E C T R O P H O T O M E T E R
11
Sample Preview Method
The instrument has three sample preview methods:
Sample Door
This method enables you to open the sample door and look at the sample position in the viewport.
You can make manual adjustments if necessary to obtain optimum sample alignment. Refer to the
instructions on using the Sample Door Preview Method.
External Video Monitor Feature
This method requires an external preview video monitor. The monitor is plugged into the video
connector located on the back of the instrument. After the monitor is connected, the
Preview
button on the spectrophotometer can be pushed to turn live preview video on and off. Video
on/off is the default functionality of the
Preview
button when using a reflectance measurement
mode. However, your software application may assign a different behavior to the button for
various reasons.
Live Preview Video in PC Software Application
This method requires a software application, such as Color iQC to provide a means to view video
within the application. Follow the instructions in your application to preview video for sample
alignment.
Note:
DO NOT power down the instrument into “Standby” mode when preview video is actively
streaming in the application. Doing so may result in an unstable operation condition.
Front Panel
The front panel is used to initiate measurements and also indicate a variety of instrument
conditions, such as calibration status, measurement status, etc. Below is description for each
indicator and button.
Specular Included
Specular Excluded
Lens
Standard button and indicator
Preview button and indicator
Trial button and indicator
Calibration
Transform
NetProfiler
Total Transmission
Direct Transmission
Haze
D65 Illuminate
UV Excluded
400, 420, 460 nm
cut filters
Custom UV (user defined)
Power “standby” button and indicator
Aperture plate