OPERATING THE DENALI 600 PATIENT LIFT
SAFETY FEATURES
EMERGENCY STOP BUTTON
Push in the red button until it locks to stop all
functions.
Turn clockwise to release and restore all functions.
POWERED EMERGENCY LOWERING OF SLING BAR
Push in the black button to lower the sling bar when
the lift control handset does not function.
MANUAL LOWERING
Turn the handle counter-clockwise to lower the
sling bar when the lift control handset or the
POWERED EMERGENCY LOWERING does not
function.
Slack in the sling bar strap
All lift arm and sling bar movement functions are
disabled if excessive slack is present in the sling bar
strap.
Apply tension to the sling bar strap to restore all
function.
Twist in the sling bar strap
The lift arm can only be raised and the sling bar can
only be lowered if the sling bar strap is twisted.
Straighten the twisted strap to restore all function.
Battery charging inhibits driving
function
The lift cannot be driven if the battery charger is
connected to the battery.
Disconnect the charger cable from the battery to
restore function.