NOx-N
What to do if...
4 - 1
ba43110e05
01/2011
4
What to do if...
Different types of messages can appear on the display:
z
Error message
Malfunction, TresCon is in the Stop mode
z
Informative message
Normal operation, consumables declining
z
Service message
Normal operation, interval for maintenance activities has expired
Note
If errors or instructive messages occur that are not listed here, further
information is given in the section W
HAT
TO
DO
IF
... of the T
ERMINAL
volume of this operating manual.
4.1
Error messages
When an error occurs, an appropriate short text appears on the display
and the analyzer module switches into the
Stop
mode. You must
employ the appropriate measures depending on the error message to
clear the fault.
The following overview contains all the possible messages and the
measures required to clear the fault.
Cause Action
Dist. H2O cannot be
transported, negative
pressure in the hose
connector.
Check the tubes for breaks and foreign
bodies and replace them as necessary
(including the tube in the container).
Check whether air can pass through the
vent in the container lid.
Cause Action
No more dist. H2O in the
container.
Replace the container with a full one.
Start the
Replenish H2O dist
service.
Dist. H2O clogged
Dist. H2O empty