wtw TetraCon 925/LV Operating Manual Download Page 6

Measurement / Operation

TetraCon

®

 925/LV

6

ba77131e02

12/2015

1.2

Recommended fields of application

Applications in the water lab

Measurements in small sample volumes and low depths of immersion. 

Titration

2

Measurement / Operation

2.1

Commissioning

Scope of delivery

Conductivity measuring cellTetraCon

®

 925/LV

Operating manual

Making the sensor

ready to measure

TetraCon

®

 925/LV

Connect the sensor to the meter. The sensor is 
immediately ready to measure.

TetraCon

®

 925/LV-P

Connect the sensor to a free IDS sensor plug-in 
position of the multi parameter probe or to an IDS 
connector of the meter. To open and close the IDS 
plug connection please follow the section 2.2 
O

PENING

 

AND

 

CLOSING

 

THE

 IDS 

PLUG

 

CONNECTION

  

(T

ETRA

C

ON®

 925/LV-P). 

The sensor is immediately ready to measure.

Connection cables in different lengths to connect 
the TetraCon

®

 925/LV-P sensor to the meter are 

listed in chapter 6 W

EAR

 

PARTS

 

AND

 

ACCESSORIES

.

Summary of Contents for TetraCon 925/LV

Page 1: ...OPERATING MANUAL ba77131e02 12 2015 TetraCon 925 LV P CONDUCTIVITY MEASURING CELL TetraCon 925 LV TetraCon 925 LV P...

Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...f application 6 2 Measurement Operation 6 2 1 Commissioning 6 2 2 Opening and closing the IDS plug connection TetraCon 925 LV P 7 3 Cleaning 8 4 What to do if 8 5 Technical data 9 5 1 General data 9 5...

Page 4: ...Contents TetraCon 925 LV 4 ba77131e02 12 2015...

Page 5: ...r measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct cell constant is automatically used if the sensor is operated with several meters...

Page 6: ...re TetraCon 925 LV Connect the sensor to the meter The sensor is immediately ready to measure TetraCon 925 LV P Connect the sensor to a free IDS sensor plug in position of the multi parameter probe or...

Page 7: ...the locking device step 1 Use your thumb and index finger to press the clips of the connector togeth er and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plu...

Page 8: ...cell as electronic waste 4 What to do if Contamination Cleaning procedure Lime sediments Immerse in acetic acid for 5 minutes vol ume share 10 Fat oil Clean with warm water that contains wash ing up...

Page 9: ...0 k at 25 C 77 F Dimensions in mm 16 16 120 0 120 0 176 0 222 8 15 3 15 3 21 7 21 7 TetraCon 925 LV TetraCon 925 LV P Weights TetraCon 925 LV 60 g without cable TetraCon 925 LV P 65 g Materials Shaft...

Page 10: ...ets the requirements according to article 3 3 of the directive 97 23 EC pressure equipment directive Measurement conditions Conductivity measuring range 1 S cm 2 S cm Temperature range 5 70 C 100 C 23...

Page 11: ...199 9 200 1999 0 1 1 mS cm 2 00 19 99 20 0 199 9 200 2000 0 01 0 1 1 resistivity Ohm cm 0 50 19 99 20 0 199 9 200 1999 0 01 0 1 1 resistivity kOhm cm 2 00 19 99 20 0 199 9 200 1999 0 01 0 1 1 resistiv...

Page 12: ...ection cable 10 m AS IDS 10 903 853 IDS connection cable 15 m AS IDS 15 903 854 IDS connection cable 20 m AS IDS 20 903 855 IDS connection cable 25 m AS IDS 25 903 856 IDS connection cable 40 m AS IDS...

Page 13: ......