16 RS 232 C interface
Meaning of the special characters:
CSPK [Lmin Lmax]
Measurement and transmission of all measured values of the spectrum
[Lmin Lmax] = wavelength range
CEME
Input user-defined methods (see chapter
)
REME [MMM]
Output user-defined methods
[MMM] = method number
CCLR [MMM]
Erase user-defined methods
[MMM] = method number
The error message, Invalid command ap-
pears if commands are unknown or cannot
be carried out (e.
g. if optional parameters do
not agree with the cell coding). Optional pa-
rameters [MMM] and [LLL] need only be in-
put for uncoded cells.
Character
Meaning
3
consecutive number (not required for interface commands CMES, CEXT and
CBLA)
5
method designation
6
ID number
17
date and time
4
special character
9
measured value
10
unit
12
citation form
4
AQA I. D. (AQA1/AQA2)
4
dilution factor
Data fields are separated by spaces.
Character set: IBM, code page 437
Character
Meaning
!
Measuring with blank value (concentration) or reference absorbance (absor-
bance)
t/T
Measuring with turbidity correction / with high turbidity
*
Measured value outside the measuring range
Q
AQA measurement
Command
Function
16.3
Output format of measured values
Summary of Contents for PhotoLab Spektral
Page 2: ......
Page 32: ......
Page 38: ...8 List of contents...
Page 128: ...98 13 User defined methods...
Page 132: ...102 14 Photometer setup...
Page 140: ...110 16 RS 232 C interface...
Page 144: ...114 18 Exchanging the Memory Card...
Page 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Page 153: ...123 21 Appendix...
Page 154: ...124 21 Appendix...
Page 160: ......
Page 161: ......