Configuring
pH 330i/340i
12
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
l
Switch off the measuring instrument.
l
While pressing the <M> key, press <
>.
l
Select the required Baud rate with <
▲
> <
▼
>.
Selection: 1200, 2400, 4800, 9600 Baud.
l
Then press the <RUN/ENTER> key.
l
Select the required display during the pH calibration with <
▲
> <
▼
>.
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l
Then press the <RUN/ENTER> key.
l
Set the required calibration interval with <
▲
> <
▼
>.
Selection: 1... 7... 999 d.
l
Then press the <RUN/ENTER> key.
l
Select the date and time step-by-step with <
▲
> <
▼
>. In between, press
the <RUN/ENTER> key each time.
l
When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.
Baud
0
4 0
8
L
C A
p
i s
d
/pH
7
3
nt
pH
Time
I
d
Day.Month
4
0 0
9