pH 315i
Operation
15
4.2.1
Measuring the pH value
Changing the
pH resolution
The measuring instrument shows the pH measured value with a reso-
lution of 0.01 or 0.001 (default setting is 0.01). To change over the res-
olution, press the <M> key while pressing the <RUN/ENTER> key.
AutoRead AR
(drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
For identical measurement conditions, the following criteria apply:
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
1
Perform the preparatory activities according to section 4.2.
2
Immerse the pH electrode into the test sample.
3
Press the <M> key until pH appears on the status display. The
pH value appears on the display.
2
%
.
1
S
Time
Baud
Day.Month No.
Ident
Year
LoBat
CalError
TempError
ARng AR
RCL
AutoCal DIN
AutoCal TEC ConCal
Auto Store
°F
mg/L
mV/pH
mol/L
% ppm
U
ISE
∆
8
4
2
99
6
°C
pH
TP
Reproducibility
Response time
Better than 0.02
> 30 seconds
1
Call up the pH measuring mode with <M>.
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
The AR display indicator flashes until a stable signal is
reached.
4
If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5
To terminate AutoRead: Press the <AR> key.