Switch the
AutoTemp
function on or off (only in operating modes
BOD Standard
and
BOD Special
):
9.4.8
Switch-off interval
The time interval after the last time a key is pressed can be set here. After
this time interval expires, the controller switches itself off to save energy.
Possible settings: 5 to 15 minutes. Default setting: 5 minutes.
BOD measuring
time
Duration of the adapta-
tion phase
Duration of the test phase
0.5 to 23 hours
Since the measurement times are very short, the sys-
tem always suppresses the
AutoTemp
phase here
even if the AutoTemp function is switched on in the set-
tings.
1, 2, 3, 4, 5 days
14, 28, 42, 56, 70 minutes Max. 28, 56, 84, 112, 140
minutes
6 to 99 days
in a 1 day pattern
70 minutes
Max. 140 minutes
SETTINGS
menu
Use
<
><
>
to select the
-
AutoTemp
menu item.
Preselected:
- AutoTemp
On
To change it:
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to select either
-
AutoTemp
On
or
Off
.
Confirm with
<RUN/ENTER>
.
- Back
(
to the
AutoTemp
menu)
RUN/ENTER
In
BOD Routine
mode the switch-off interval
is permanently set to
5 min and cannot be changed.
SETTINGS
menu:
Use
<
><
>
to select the
-
Switch-off interv.
menu item.
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......