9.4.6
Data storage
Here, you can set whether the controller should automatically erase the
oldest finished samples if the memory is full to create space for new
measurement data (setting
auto
).
With the
manual
setting, the message "
Memory lack!
Erase finished
sample/s!
is displayed if the memory is full (see section 12 W
9.4.7
AutoTemp
The
AutoTemp
function controls the automatic start of the measure-
ment after checking the temperature adaptation.
The pretemperature regulation to the precise incubator temperature is
recommended but not essential. Recommendation: e.g. regulate the
temperature of the sample for BOD5 measurements from 15 °C up to
20 °C.
You can tightly close the sample bottle with the measuring head imme-
diately and start the measurement. The
AutoTemp
function then auto-
matically triggers the start of the actual measuring after checking the
temperature adaptation. The measuring time of the
AutoTemp
phase
(adaptation phase plus the test phase) is added to the sample
measuring time selected in the settings.
Pretemperature regula-
tion of the sample
With the
AutoTemp
function switched on and, adhering to the recom-
mendation according to the table, the error quota that results from the
temperature adaptation of the sample to the incubator temperature,
Back up your data regularly!
SETTINGS
menu:
Use
<
><
>
to select the
-
Memory
menu item.
Preselected:
MEMORY
- Erase
auto
To change it:
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to select either
auto
or
manual
.
Confirm with
<RUN/ENTER>
.
- Back
(
return to the
SETTINGS
menu)
RUN/ENTER
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......