Data storage
Oxi 3315
40
ba75944e02
03/2012
confirm with
<ENTER>
. The meter switches to the measured value display.
The active automatic storage function can be recognized by the progress bar
in the status line. The progress bar indicates the remaining storage duration.
Terminating the
automatic storage
function prema-
turely
Proceed as follows to switch off the automatic storage function before the
adjusted storage duration has expired:
1 Remaining storage duration
2 Graphical display of the storage duration
If the automatic data storage function is activated, only the follow-
ing keys are active
:
<M>
,
<STO_>
and
<On/Off>
. The other keys
and the automatic switch-off function are deactivated.
0d03h00min
1
2
01.09.2011
08:00
1.
Press the
<STO_>
key.
The following window appears.
2.
Using
<
><
>
, select
yes
and confirm with
<ENTER>
.
The meter switches to the measured value display.
The automatic storage function is terminated.
Stop automatic storage?
yes
no
Warning
01.09.2011
08:00