Configuring
Oxi 330i/340i
10
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
l
Switch off the measuring instrument.
l
While pressing the <M> key, press <
>.
l
Select the required Baud rate with <
▲
> <
▼
>.
Selection: 1200, 2400, 4800, 9600 Baud.
l
Then press the <RUN/ENTER> key.
l
The air pressure is only displayed if an oxygen sensor is connected. Oth-
erwise, "---" appears.
l
Continue with the <RUN/ENTER> key.
l
Select the required calibration interval for oxygen measurements with
<
▲
> <
▼
>.
Selection: 1... 14... 999 d.
l
Then press the <RUN/ENTER> key.
l
Switch the automatic measuring range selection AutoRange on/off with
<
▲
> <
▼
>.
Selection: YES (On), no (Off).
l
Then press the <RUN/ENTER> key.
l
Select the date and time step-by-step with <
▲
> <
▼
>. In between, press
the <RUN/ENTER> key each time.
l
When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.