Conductivity
Multi 9310
56
ba75937e03
12/2011
8
Conductivity
8.1
Measuring
8.1.1 Measuring the conductivity
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
z
Conductivity [
µ
S/cm] / [mS/cm]
z
Resistivity [
Ω
·cm] / [k
Ω
·cm] / [M
Ω
·cm]
z
Salinity SaL [ ]
z
Total dissolved solids TDS [mg/l] / [g/l]
The factor to calculate the total dissolved solids is set to 1.00 in the factory. You
can adjust this factor to meet your requirements in the range 0.40 ... 1.00. The
factor is set in the menu for the parameter, TDS.
Stability control
(AutoRead)
& HOLD function
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
z
as soon as the measured value is outside the stability range
z
when the automatic
Stability control
is switched off.
You can start the
Stability control
manually at any time, irrespective of the set-
1.
Connect the conductivity sensor to the meter.
The conductivity measuring window is displayed.
The measuring cell and cell constant for the connected IDS conductivity
sensor are automatically taken over.
2.
If necessary, select the measured parameter
ϰ
with
<M>
.
3.
Immerse the conductivity sensor in the test sample.
Info
01.07.2011
08:00
USB output