pH value
Multi 9310
28
ba75937e03
12/2011
5
pH value
5.1
Measuring
5.1.1 Measuring the pH value
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
Stability control
(AutoRead)
& HOLD function
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
z
as soon as the measured value is outside the stability range
z
when the automatic
Stability control
is switched off.
You can start the
Stability control
manually at any time, irrespective of the set-
ting for automatic
Stability control
, page 71) in the
System
menu.
1.
Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2.
If necessary, select the measured parameter with
<M>
.
3.
Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.
4.
If necessary, calibrate or check the IDS pH sensor.
5.
Immerse the IDS pH sensor in the test sample.
Info
01.07.2011
08:00
USB output
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
The HOLD function is active.