MIQ/CR3; DIQ/CR3
Technical data
19
ba64107d13
03/2018
6.2
MIQ/CR3; DIQ/CR3
Electrical data
Terminal
connections
Nominal voltage
Max. 24 VDC via the IQ S
ENSOR
N
ET
(for
details, see the T
ECHNICAL
DATA
chapter of
the IQ S
ENSOR
N
ET
system operating
manual)
Power consumption
MIQ/CR3: 3 W
DIQ/CR3: 2,3 W
Protective class
II
Overvoltage
category
II
IQ S
ENSOR
N
ET
connections
2
Additional connectable SENSORNET
terminator (terminating resistor)
Number of relay
contacts
3
Number of current
outputs
3
Terminal type
Screw-type terminal strip, accessible by
opening the lid
Terminal ranges
Solid wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
Flexible wires:
0.2 ... 2.5 mm
2
Line cross-section of
cables carrying mains
voltage
Europe:
1.5 ... 4.0 mm
2
USA:
AWG 14 ... 12
Cable feeds
4 cable glands M16 x 1.5 on the underside
of the module