Commissioning
IFL 70x IQ
3 - 10
ba75990e01
10/2012
Setting
Selection/values
Explanation
Method
The rising side of the echo is evaluated for
the measured value determination. One of
two methods can be selected for this.
Rel. threshold
The measured value is equivalent to the
point where the intensity of the echo
reaches the adjusted relative threshold.
The value relates to the maximum of the
echo intensity (100 %):
Maximum gradient
The measured value is equivalent to the
point with the maximum gradient:
Rel. threshold value
25 ...
75
... 100 %
Threshold value for evaluation according
to the
Rel. threshold
method.
100 %
Depth
Measured value
25 %
Relative
threshold
Depth
Measured value
Maximum
gradient