Within the proportional range an increase of the concentration by 1 mg/
l causes a reduction of the output current by 1 mA. If the measured
concentration is 17 mg/l, for example, 11 mA is output:
The highest concentration at which the controller still works in the
proportional range is 8 mg/l (corresponding to Imin = 8 mA) and the
lowest is 2 mg/l (corresponding to Imax = 14 mA).
Settings
(
)
mg/l
mg/l
mg/l
mA
1
mA
12
I
Controller
5
4
−
⋅
+
=
(
)
mA
11
mg/l
1
mg/l
mA
1
mA
12
I
Controller
=
−
⋅
+
=
Setting
Selection/Values
Explanation
Measured variable
Main variable
Adjoining variable
Main variable
designates the actual
measured parameter
of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).
Nominal value
Within the measuring
range (sensor-
dependent)
Nominal value the
measured value is
regulated to
Xp
5 ... 100 %
-5 ... -100 %
% of the measuring
range
Proportional range of
the controller.
Negative values
result in a positive
characteristic curve.
Imin
0 ... 20 mA
Lower current
limitation *
Imax
0 ... 20 mA
Upper current
limitation *
* Note:
Spacing between
Imin and Imax:
at least 5 mA
Summary of Contents for DIQ/S 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Page 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Page 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Page 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Page 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Page 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Page 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Page 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Page 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Page 151: ......