Operation
4 - 3
ba75504e07
04/2012
Select the
PROPERTIES
menu with
g
. With this menu you can put
sensors in the maintenance condition and prompt important data
(calibration data, error messages, operating states, software
versions, etc.).
Open the
SETTINGS
menu with
s
.
Special
sensor conditions
The following displays inform you of special states of sensors
Error and info symbol
If the info symbol
i
or error symbol
h
appears in the header, error
messages from the sensors or error conditions of the system have
occurred. Error messages from the sensors are entered in the log book.
The log book is described in detail in section 4.4.3. Error conditions of
the system can e.g. be an insufficient operational voltage or
malfunctions of the communication and are displayed in the
PROPERTIES /
SYSTEM STATUS
menu (see section 4.4.1).
Init
Sensor is being initialized
during commissioning or
if a new IQ sensor is recognized that is not yet
giving measured values
----
Sensor inactive or inadmissible operating conditions
Cal
Sensor is being calibrated
Clean
Cleaning procedure active
Error
No communication with the sensor due to defective
connection
OFL
Measuring range undercut or exceeded (overflow)
Display
flashes
Sensor in maintenance condition
Summary of Contents for DIQ/S 182-MOD
Page 6: ...Contents System 182 0 4 ba75504e07 04 2012...
Page 10: ...Overview System 182 1 4 ba75504e07 04 2012...
Page 14: ...Safety instructions System 182 2 4 ba75504e07 04 2012...
Page 78: ...Operation System 182 4 18 ba75504e07 04 2012...
Page 84: ...The SETTINGS menu System 182 5 6 ba75504e07 04 2012...
Page 110: ...Relay outputs System 182 6 26 ba75504e07 04 2012...
Page 126: ...Maintenance and cleaning System 182 8 2 ba75504e07 04 2012...
Page 132: ...What to do if System 182 9 6 ba75504e07 04 2012...
Page 144: ...Accessories and options System 182 11 2 ba75504e07 04 2012...
Page 148: ...Appendix store separately if required System 182 13 2 ba75504e07 04 2012...
Page 149: ......