Operation
DIQ/S 181
46
ba77083e01
02/2015
4.2
Measured value and status display
With the
<M>
key, you can change from any operating situation to the
measured value display or status display last selected. Entries that are
not completed are ignored while doing so.
Pressing
<M>
once again cyclically switches between the further dis-
play options.
Example:
Display options
Operating notes:
With
<C>
you can start a calibration procedure for the sensor. Start-
ing calibration from the display of outputs is not possible.
Using
<OK>
, call up the menu
PROPERTIES
. This menu can be
used to put the sensor in the maintenance condition and prompt for
important data (calibration data, error messages, operating states,
software versions etc.).
Using
<S>
, open the
SETTINGS
menu.
Special
sensor conditions
The following displays inform you of special states of the sensor:
Display of measured values.
Main and secondary measured
parameter.
Display of the outputs.
The current states of the relays
and current outputs and the date
and time are displayed.
Init
Sensor is being initialized
during commissioning or
if a new sensor is recognized that is not yet giving
measured values
----
Sensor inactive or inadmissible operating conditions
Cal
Sensor is being calibrated
Clean
Cleaning procedure active
Error
No communication with the sensor due to defective
connection
OFL
Measuring range undercut or exceeded (overflow)
Display
flashes
Sensor in maintenance condition
Summary of Contents for DIQ/S 181
Page 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Page 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Page 103: ......