Conductivity
Cond 7110
22
ba75927e03
04/2012
3.
Confirm the selection of (
Use Cell
) with
<ENTER>
or
<CAL__>
.
The cell constant that was set last is displayed.
4.
Select a cell constant in a suitable range with
<CAL>
.
•
Range 0.250 ... 2.500 cm-1
•
Range 0.090 ... 0.110 cm-1
•
Permanent, 0.010 cm
-1
5.
Set the exact cell constant within the range with
<
><
>
as necessary.
6.
Use
<ENTER>
to finish the calibration.
The adjusted cell constant is used.
5.2.6
Calibration data
Displaying the calibration data
1.
Show the calibration data in the measured value display with
<CAL__>
.
The calibrated cell constant is displayed.
L
C L
E
E
S
U
1/cm
6
0
7
4
L
A
C
Cal