Settings
Cond 3310 IDS
30
ba77051e04
07/2017
6
Settings
6.1
Cond measurement settings
6.1.1 Settings for IDS conductivity sensors
Settings
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<ENTER>
key. After completing the settings,
switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
TetraCon 925
,
LR 925/01
).
Default settings are printed in
bold
.
Setting
menuTetraCon 925
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the last calibration
records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data
memory to the interface
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the IDS
conductivity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Type
Used measuring cell
Cal
Measuring cells whose cell con-
stant is determined by calibra-
tion in the KCL control standard
solution.
Calibration range:
0.450 ... 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man
Freely (manually) adjustable cell
constant in the range 0.450 ...
0.500 cm
-1.