Conductivity
Cond 3310 IDS
26
ba77051e04
07/2017
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
5.2
Temperature compensation
The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C or 25 °C. It appears on the display as
Tr20
or
Tr25
.
You can select one of the following temperature compensation methods:
Nonlinear temperature compensation
(
nLF
) according to EN 27 888
Linear temperature compensation
(lin)
with adjustable coefficients of
0.000 ... 3.000 %
/
K
No temperature compensation (off)
Application tips
Select the following temperature compensations given in the table according to
the respective test sample:
5.3
Calibration
5.3.1 Why calibrate?
Aging slightly changes the cell constant, e.g. due to coatings. As a result, an
inexact measured value is displayed. The original characteristics of the cell can
often be restored by cleaning the cell. Calibration determines the current value
of the cell constant and stores this value in the meter.
Thus, you should calibrate at regular intervals.
The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 6.1.1 S
Test sample
Temperature compensation
Display
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
lin
Set linear
temperature coefficient
0.000 ... 10.000 %/K
lin
Salinity (seawater)
Automatic
nLF
according to
IOT (International Oceano-
graphic Tables)
Sal
,
nLF
Summary of Contents for Cond 3310 IDS
Page 6: ...Contents Cond 3310 IDS 6 ba77051e04 07 2017...
Page 56: ...Index Cond 3310 IDS 56 ba77051e04 07 2017...