AmmoLyt

®

NHA/AT

2

Caution

The membrane and junction of the electrode must never 

!

dry up (observe notes on storage)

!

be damaged

!

be brought into contact with grease.

Note

After installing the electrode tap your finger against the combination 
electrode to remove any air bubbles from the back of the membrane. 
When doing so hold the electrode in an upright position (membrane 
downward).

Cleaning

In built-in condition:
Splash the AmmoLyt

®

NHA/AT electrode with water. Remove coarse 

contamination using a soft toothbrush under running water.

Note

For the Technical Data of the electrode, refer to the operating manual 
of the AmmoLyt

®

NHA reference electrode.

Instructions on how to install the electrode in the AmmoLyt

®

700 IQ 

sensor are given in the operating manual of the AmmoLyt

®

700 IQ sen-

sor.

2

Screw the AmmoLyt

®

NHA/AT exchange electrode into the 

AmmoLyt

®

NHA reference electrode without leaving any gap. 

A flat wrench (wrench size 13 mm) or the protective hood of the 
sensor is suitable as a wrench (see following figure).

Protection hood