Ωmega-Tip-Z
World Precision Instruments
5
Battery Replacement
Batteries will rarely require replacement. Replace batteries annually or when the
Lo-
Batt
indicator on the LCD digital meter is displayed.
To replace batteries, remove four screws on the bottom of the instrument case.
Withdraw the front panel and remove the old battery cells. Install six AA alkaline cells
as the polarity markings indicate. Close the case and test the instrument using the
Brief Test described above.
To conserve batteries, remember to turn power off when the instrument is not in use.
SPECIFICATIONS
INPUT RESISTANCE....................................................................................................................... 10
12
Ω
, typical
INPUT LEAKAGE CURRENT
............................................................................................................1 pA, typical
MAX. INPUT VOLTAGE SWING ...................................................................................................................± 2 V
VOLTMETER
Range .........................................................................................................................0 to ±2000 mV DC
Accuracy ......................................................................
± 0.1% of reading, ± 1 least significant digit
Resolution ..................................................................................................................... 1 mV (3.5 digits)
OHMMETER
............................................................................................................................................................
Range .....................................................................................................0 to 20 M
Ω
@ 500 Hz (metal)
................................................................
0 to 200 M
Ω
or 0 to 2000 M
Ω
@ 12.5 Hz (micropipette)
Measurement .....................................................................................10 nA and 1 nA (micropipette)
..............................................................................................................................................
10 nA (metal)
Resolution ........................................................................................................................................
10 K
Ω
Accuracy ...........................................................................................................................................± 20%
PROBE ............................................................................................................1.2 cm diameter
×
3.2 cm long
PROBE HANDLE ......................................................................................... 4.7 mm diameter
×
8.9 cm long
PROBE CABLE ...............................................................................................................................................1.5 m
MAIN HOUSING............................................................................................7
×
4
×
2 in. (18
×
10
×
5 cm)
POWER ........................................................................................................6 AA 1.5 V alkaline cells, supplied
SHIPPING WEIGHT ..........................................................................................................................4 lb. (1.8 kg)