Calibrating the Sensor

Preparing Na

2

S Stock Solution

1.  Use a 100mL volumetric flask. Dissolve 5mg EDTA in 100mL distilled water.
2  Purge the solution vigorously with argon gas for 15 minutes. 
3.  Weigh 24.0mg of sodium sulfide (Na

2

S .9 H

2

O) (reagent plus, 99.99+% from Sigma) 

and dissolve it in the solution under argon atmospheres. 

4.  Seal the flask with a rubber stopper. The solution is 1.0mM Na

2

S. Store the solu-

tion in a dark bottle and refrigerate at 2–8°C.

Calibration Procedure

The sensor should be immersed in PBS and plugged into the free radical analyzer. Set 

the range to 100nA and the poise to H

2

S (150mV). Once the sensor is polarized, It can 

be calibrated. The following example describes the fundamental concepts behind a 

standard calibration protocol. Known concentrations of H

2

S are generated in PBS by 

adding a known volume of a the Na

2

S stock solution.

1.  Place a 20mL vial of PBS on the magnetic stirring plate with the small stir bar still 

inside. Turn on the stirrer so that the bar is stirring at a moderate rate.

 

NOTE

: This rate of the stirrer should NOT be modified during calibration.

 NOTE

: The calibration should be carried out at the same temperature at which 

the experimental measurements are to be made. This can be accomplished by 

placing the vial and stand in a water bath at the appropriate temperature and 

allowing the temperature of the solution in the vial to equilibrate with the water 

bath. 

2.  Immerse the 

ISO-H2S-2

 sensor in the solution and secure it in an electrode 

holder such as WPI’s 

Pro-Guide

 (WPI #

47510

47520

47530

47540

) or a  micro-

manipulator. The sensor tip should be immersed about 2–3mm into the solution. 

It should not contact the stir bar, which could damage the membrane. 

3.  Connect the H

2

S sensor into a channel of a free radical analyzer. Set the poise 

voltage to +150mV. For the 

Apollo 4000

, this is accomplished in the software. 

Use the custom option to manually set the voltage to 150mV. For the 

Apollo 

1000

 select poise from the main menu and set it to “internal”. Then, adjust the 

poise voltage using a small screwdriver to turn the “poise adjust” potentiometer 

on the front panel. For the 

TBR4100

, simply set the “probe select” control on the 

front panel to H

2

S. Allow the sensor to polarize for 12 hours.

 The 

quiescent baseline current is an indicator of the health of the sensor. A cor-

rectly operating sensor has a baseline current of < 75,000pA. 

4.  To the vial containing 20mL PBS, sequentially inject three aliquots of Na

2

S solution 

(20µL, 40µL and 80µL). The current output increases after each addition and then 

plateaus. As soon as it reaches a plateau, inject the next aliquot. 

Summary of Contents for ISO-H2S-2

Page 1: ...Instrumenting scienti c ideas WORLD PRECISION INSTRUMENTS ISO H2S 2 Hydrogen Sulfide Sensor for use with WPI Analyzers Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 030316...

Page 2: ...Solution 95611 7325 100042 7326 100084 Start up Kit 95699 5435 600011 5377 600015 Higher detection limit available on request call for custom pricing Real time detection using electrochemical microsen...

Page 3: ...ENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 Notes and Warnings 1 Parts List 2 Unpacking 2 INSTRUMENT DESCRIPTION 2 OPERATING INSTRUCTIONS 3 Polarizing the Sensor 3 Calibrating the Sensor 4 Preparing Na2 S...

Page 4: ...ii World Precision Instruments...

Page 5: ...able protective stainless steel sleeve WPI 600016 which must contain fresh electrolyte WPI 100084 and is separated from the external environment by a gas permeable polymeric membrane covering the end...

Page 6: ...f any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Departmen...

Page 7: ...ne sleeves are available in packages of 4 WPI 600016 The ISO H2S start up kit WPI 600015 also includes replacement membrane sleeves along with all the accessories to fill them properly with electrolyt...

Page 8: ...al measurements are to be made This can be accomplished by placing the vial and stand in a water bath at the appropriate temperature and allowing the temperature of the solution in the vial to equilib...

Page 9: ...f to 20mL of PBS The concentration of H2 S produced can be calculated as follows M1 V1 M2 V2 where M molarity and V volume 1st Aliquot 20 L 20mL 20 020 L 1000 M 20 L M2 20 020 L 20 000 M 20 020 0 999...

Page 10: ...f this curve indicates the sensitivity of the sensor Once the sensitivity of the probe has been determined the sensor is ready to use experimentally In the above example the sensitivity was 1 98pA nM...

Page 11: ...brane Cleaning the Membrane The membrane sleeve itself requires little maintenance The primary concern is to avoid damage to the membrane and to keep it as clean as possible After each use clean the m...

Page 12: ...3 and 4 6 Remove the locking cap from the old used sleeve and gently slide it onto the new replacement sleeve Additional membrane sleeves WPI 600016 may be pur chased separately 7 Dip the internal ele...

Page 13: ...ACCESSORIES Part Number Description 5399 T Adapter Kit pkg of 3 600015 ISO H2S 2 Startup Kit 600016 Replacement Sleeves pkg of 4 100084 ISO H2S 2 Electrolyte Filling Solution 10mL 7363 4 Enzol Enzymat...

Page 14: ...polarizing solu tion may be contaminated Prepare fresh polarizing solution Use 0 1M PBS only See Replacing the Mem brane Sleeve on page 8 External electrical interferences may be the problem Identify...

Page 15: ...Outside Diameter 2mm Response Time 10 sec Detection LImit Range 5nM Nominal Sensitivity New sensor 1 pA nM Poise Voltage 150mV Typical Quiescent Baseline Current 25 C 75000pA Acceptable Baseline Rang...

Page 16: ...12 World Precision Instruments...

Page 17: ...r regression 6 locking cap 1 3 M membrane 1 3 membrane sleeve replacement 8 MF28G67 5 8 9 MicroFil 8 O organic matter 7 P parts list 2 PBS 1 3 4 5 10 11 polarization 3 polarize 1 probe handle 3 Pro Gu...

Page 18: ...14 World Precision Instruments DECLARATION OF CONFORMITY...

Page 19: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Page 20: ...com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 29a No8 Dongfang Rd Pudong District Shanghai 200120...

Reviews: