6
World Precision Instruments
SPECIFICATIONS
This unit conforms to the following specifications:
Input Resistance ...............................................>10
12
Ohms, common mode and
differential
Input Leakage Current .................................................................................................
50 pA, max.
Amplification
.......................................................................................
100×, 1000×, 10,000× (AC)
........................................................................................................................
10×, 100×, 1000× (DC)
Common Mode Rejection Ratio (CMRR)
..........................................
100 dB typ. @ 50/60 Hz,
..............................................................................................................................
80 dB typ. @ 1 KHz
Equivalent Noise Signal Input (ENSI)
..............................................................................................
AC amplifier (typ.)
...............................................................
0.4 µV rms (2.0 µV p/p), 0.1-100 Hz
................................................................................................
2.0 µV rms (10 µVp/p), 1 Hz-10 kHz
DC Amplifier
.............................................................................
6.0 µV rms (30 µV p/p), 3-10 kHz
Bandwidth Filter Settings
..................................................................................................................
AC ................................................................................................
low frequency 0.1, 1, 10, 300 Hz
AC and DC ..........................................................................
high frequency 100 Hz, 1, 3, 10 KHz
Position .................................................................................................................................... 250mV
Output Voltage Swing ............................................................................................................... ± 8 V
Output Resistance ...........................................................................................................470 Ohms
Battery Test.. ...............................................................................................1 second audible tone
Calibrator Signal .........................................
10 Hz square wave, amplitude 100 µV (at 1000×
..................................................................................
and 10,000×), 1 mV (at 100×) and 10× DC).
Power .......................................................................................... 2 9V alkaline batteries, supplied
Input Cables ........ 1 shielded cable with metal electrode connectors are provided and 1
shielded cable with bare wires.
APPENDIX A DAM50 WIRING CONFIGURATIONS
PART NUMBER
DESCRIPTION
M3301EH
Replacement Electrode Holder
M3301R
Manual Micromanipulator
MD4R
Dual Tool-Holder Micromanipulator
TM31A10
Tungsten Metal Electrodes Profile A, 76 mm, 1 micron
RC1T
Reference Cells
M10
M10 Magnetic Stand
5052
Steel Base Plate, 10 lbs
3294
Ground Wire for DAM80 Probe
EP2
Ag/AgCl Electrode
2505
Electrode Handle 6.3 mm
5444
Electrode Handle 4.8 mm
13620
Low-noise cable for Microelectrode Holder
5489
Optional Probe for DAM50
MEH7W-10
Microelectrode Holder
Summary of Contents for DAM50