Plug in the unit and turn the 87 switch to position I
11.
Press the Vacuum key to start the vacuum pump and keep evacuating for about 30 minutes
12.
When the vacuum operation is complete, close the 82 valve and disconnect the valve ref. 56 from the
88 service connection
13.
Remount the protective cap on the 88 connection, the plastic cover and the rear door
14.
When the filter replacement procedure is completed, turn the unit on by turning the 87 switch to
position I
15.
When the message “Replace filter” appears on the display, if you press the
key, you enter the
procedure for filter replacement. It is necessary to type the serial number of the filter and then confirm
by pressing the
key.