TANDA Color Steam Sterilizer Operation Manual

 

 

Rev-C                                                                                                                                                  Page  13  of  17 


 

 

6

Light of “END” will be lighted up when the program ends, when it releases automatically to 

the safe pressure below 0.1 bar, 

Manually press the release button to open, or rotate the handle to the door-opened position, 

the program will automatically open the exhaust valve

 

Wait for about 2-3min, then open the door

 

Press the Dry key to start the drying function, dry light flashes, when the drying indicator light   

turns to be always bright, drying process finished

 

Please take the instruments out with tools and then turn off the power switch. 

 

5.3 Drying Function Setting 

1. If you have not pressed DRY button to start the automatic drying function before you start 

the program, there will not be a drying process. After the process of exhaust, the END light is 

always  on,  the  pressure  window  shows  the  real-time  pressure,  the  temperature  window 

alternately displays the real-time temperature and "

", indicating the end of the program.   

You can open the release valve, wait for 2-3 minutes, then open the door to take out the items. 

 

2. 

Before starting the sterilization, if the DRY button is pressed in advance to start the automatic 

drying function, the program will automatically complete 20 min of drying process after the end 

of sterilization. In the drying stage, DRY lights flashes, END light flashes, and pressure window 

displays real-time pressure, temperature window alternately displays real-time temperature and 

"

". 

 

Summary of Contents for TANDA COLOR

Page 1: ...on Manual Woson Medical Instrument Co Ltd Technical Publications TANDA COLOR TANDA Color Steam Sterilizer Operation Manual ASS090 REV C Operation Documentation All rights reserved 2017 By Ningbo Jiang...

Page 2: ...11 27 Update REV C 2017 07 25 Update Please verify that you are using the latest revision of this document Information pertaining to this document is maintained on manufactory If you need to know the...

Page 3: ...ulatory requirements of the following Council Directive 93 42 EEC concerning medical devices The CE label affixed to the product testifies compliance to the Directive The location of the CE marking is...

Page 4: ...ommendation 6 2 3 Operation Risk 7 2 4 Protection Device 7 Chapter 3 Receiving and Installation 8 3 1 Check the Package 8 3 2 Unpack the Accessories 8 3 3 Installation Environment 9 Chapter 4 Descript...

Page 5: ...before attempting to use the product Keep this manual with the sterilizer at all times Periodically review the procedures for operation and safety precautions 1 2 Usage Indications Application to all...

Page 6: ...ttention to the high temperature in the chamber and the sterilizer exterior when exhausting system is running Protective Earth Indicates the protective earth grounding terminal CAUTION Dangerous volta...

Page 7: ...ordance with the instructions listed in this manual The sterilizer could not be used for liquid The sterilizer has not been designed to operate in the presence of gas or explosive vapors The trays and...

Page 8: ...finishes the cycle please keep an appropriate distance because the chamber still has rudimental steam with high temperature avoid to be scalded Every time open the door after the sterilizer finishes...

Page 9: ...equipment Front Flank Item 18 L 23 L Packing Size 660x525x495 mm 770x530x495mm N G Weight 25 28 KGs 30 33 KGs 3 2 Unpack the Accessories Open the door and take out all the accessories item listed belo...

Page 10: ...the sterilizer should be set in a level place which at least has 10 cm distance with each side 20cm of the top as followed The sterilizer should be set in a place with good ventilation The temperature...

Page 11: ...V 1800W 4A NO Item Description 1 Door Handle Open or close the door 2 Function Panel Show sterilization information for control and operation 3 Safety Valve Leak pressure automatically when over worki...

Page 12: ...ally 10 Dry lamp Flash when you choose drying process Lighted up when it ends Controlling Button Description Choose sterilizing temp 121 134 Release Open or close the release valve DRY Choose drying p...

Page 13: ...stick 4 Put the instruments which need to be sterilized in the chamber Turn on the power switch Turn the handle clockwise until it is tightened 5 Select sterilizing temperature Press the DRY key to st...

Page 14: ...1 If you have not pressed DRY button to start the automatic drying function before you start the program there will not be a drying process After the process of exhaust the END light is always on the...

Page 15: ...dry key DRY for time adjustment press the release key Release for drying time adjustment between 1 2 Drying time 1 refers to the automatic running drying time drying time 2 refers to the manual runni...

Page 16: ...Error Code List No Error Code Description 1 Er01 Sensor of the chamber defective 2 Er02 In chamber sensor defective 3 Er03 Sterilization fail high temperature 4 Er04 Sterilization fail low temperature...

Page 17: ...dix Piping and Circuit Diagram Piping Diagram RV PR EV1 TS1 EV1 PR RV TS1 PT100 AD AD EV2 EV2 Code Name Code Name RV Safety Valve EV2 Drain valve Normal Close AD Pressure Exhausting Port PR Pressure s...

Page 18: ...f 17 Circuit Diagram AC10V AC20V AC220V Transformer L N L N L 1 Heating Ring Power Socket Fuse Ground Exhausting Air Solenoid Valve Jiggle Switch Inside Temperature Sensor Outside Temperature Sensor P...

Reviews: