20
Warning:
Clean and disinfect the main unit and foot controller regularly. When subjecting the main unit and foot controller to cleaning and
disinfection ensure that the charging cable is not connected and that the charging socket is closed.
Warning:
Only the following parts can be sterilized:
Handpiece, motor and its tail cable, holder of silicone handpiece, motor disinfection stopper, motor cable-cooling pipe clamp, and
lubricant nozzle.
Concrete steps:
Step
Manual method
Automatic method
Warning
Cleaning
Separate the main unit and foot controller from
the charging base. Thoroughly wipe all visible
surfaces of the device with a disposable soft
cloth, including water bottle hooks, pedals, and
cables. And then dry them after washing.
Parts that can be cleaned automatically
are motor holders, motors and motor
tail cables.
It is recommended to use a cleaning
machine in accordance with ISO
15883-1 and refer to the instructions
of the cleaning machine.
Do not automatically clean
the main unit.
Disinfection
After any pre-cleaning, wipe all visible surfaces
of the device including water bottle hooks, pedals
and cables with a disposable soft cloth dampened
with disinfectant(at least 30 s) to ensure that all
surfaces are wet. and allow to act (contact time >
30 minutes). If large surfaces have to be treated,
use more than one wipe where necessary.
Parts that can be disinfected
automatically are motor holders,
motors and motor tail cables.
It is recommended to use a thermal
disinfector in accordance with ISO
15883-1 and refer to the instructions
of the thermal disinfector.
Before disinfection, please
plug the motor disinfection
stopper in the motor.
Drying
Dry all the cleaned and disinfected parts
thoroughly in the air indoors.
Drying is generally a part of cleaning
and disinfection process. Please
operate as per the Instruction of
washers and disinfectors.
Disposable water pipes
are for single use only and
should not be disinfected and
sterilized, and have no need
of drying.
Summary of Contents for DTE Implant-X