2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
69
Figure 2-56 Replacement of High-angle Receiving Plate
1 - M3×6 inner hex screw (×2)
2 - Fixed seat of side panel card
3 - Side basal plate
4 - M2.5×6 Cross recess head screw (2x)
5 - High-angle receiving plate
2.4.6 Common Problems and Solutions
Problem Description
Possible Cause
Solution
There is no scattergram
signal when testing the
DIFF signal of whole
blood.
⚫
The laser
driver panel is
not working.
⚫
The optical
preamplifier
panel is not
working.
Check if the laser is working.
⚫
If not, replace the laser driver panel.
⚫
If the laser is working, check the power
supply of the optical preamp panel and
replace the optical preamp panel.