5 Comprehensive Device Tuning
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
104
5
Comprehensive Device Tuning
Since the replacement and maintenance of some components can lead to changes in relevant
tuning parameters, this section introduces the tuning procedures for the parameters that may be
affected.
5.1 Position Adjustment
NOTE: The swab height needs to be reset after the replacement of the sampling assembly and
relevant parts of the sampling swab.
After the replacement of the sampling assembly or swab, perform the following tuning procedures.
1. Enter the
Service
screen and click
Position Adjustment
.
2. Click
Start
and check the distance between the bottom of the sample probe and the bottom of
the swab. If the sample probe is lower than the swab, click
Up
; if higher, click
Down
. Each click
will move the sample probe slightly. In the end, the bottom of the sample probe will be aligned
with the bottom of the swab. Then click "OK".
3. Next, click
Check
to run the initialisation of the sample probe. Check if the bottom of the sample
probe is still aligned with the bottom of the swab (see Figure 5-1). If it is, click
OK
to complete the
tuning; if not, repeat Steps 1~2.