V Quality Control
Quality control can be carried out by testing the InSight V-IA quality control cartridges.
Use InSight V-IA Veterinary Immunoassay Analyser to determine the concentration of the
test cartridges. There are 3 control level cartridges – low, mid and high.
Continue to use the analyser if the quality control result falls within the target value range
provided. If results fall outside the target value, repeat the QC with a fresh QC cartridge
and if results fall again outside the target value, stop using the analyser and contact
Woodley Equipment Company.
VI Further Product Information
VI.1 Security Classification of Medical Electrical Equipment
⚫
Type of protection against electric shock is Class I.
⚫
Pollution grade is Class 2.
⚫
Facility category (overvoltage category) is Class II.
VI.2 Contraindications
No.
VI.3 Warnings, Precautions and Limitations
Notices:
For veterinary use only.
VI.3.1 Precautions
Warnings:
⚫
To avoid electrical overload and potential fire risk, do not use a multi-socket adapter.
⚫
Use a 12V/5A power adapter and an effectively grounded outlet.
⚫
A damaged, non-original or modified power cord is a potential fire and electric shock
risk. Do not bend or roll the power cord so as to avoid a fire or electric shock.
⚫
If the analyser is damaged or has been dropped, please contact Woodley Equipment
Company.
⚫
Do not use this analyser in unstable environments such as on unlevel or vibrating
surfaces etc.
⚫
Do not place the analyser in a location where it is difficult to disconnect the device.
⚫
Water or debris should not enter the analyser. If this occurs, please contact Woodley
Equipment Company.