sETup locaTion:
1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 51 cms is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for
safety. Choose a setup location which will allow for a clearance envelope of at least
76 x 76 cms, (with the centrifuge at the center). No person or hazardous material shall
be permitted in the clearance envelope during operation. The operator time within
the envelope shall be limited to the time necessary for loading, unloading and
centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as
premature failure of the centrifuge. Choose an area which will allow unencumbered
air flow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary
for leveling the centrifuge, however, the surface should be flat and level.
e)
be sure the outlet is always within reach as the mains lead is the means of
emergency disconnection!
iniTial sETup procEdurE:
If any problems are found during the initial setup
procedure, refer to the troubleshooting section. For further assistance, contact Woodley Equipment
Company LTD. at +44 (0) 1204 669033
1. Plug the female end of the mains lead into the rear of the centrifuge. Plug the male end into
an approved electrical outlet. For electrical safety, the unit must always be properly
grounded.
2. For operator safety, the locking system is always active; requiring power and direction
from the user to disengage it (the lid also automatically unlocks at the end of a run when
it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the
‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ indicator light should
illuminate. If it does not, refer to the section on troubleshooting. The lid will remain
unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
3. Turn the lock counterclockwise and open the lid.
4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely,
refer to the section on troubleshooting.
5. Place the test tube holders (or carriers depending on your rotor type) inside the rotor and
verify that they are seated properly.
6. Close the lid. Rotate the lid knob clockwise to its complete stop position. The
’LOCKED’ indicator light should be illuminated. If it is not, make sure that the lid is
locked properly. The centrifuge will not run unless the lid is latched properly and the
’LOCKED’ indicator light is illuminated.
7. Use the up arrow button next to the RPM display to set the speed to full.
8. Initiate a test run by pressing the ‘START’ button.
9. The ‘RUNNING’ indicator light will illuminate.
10. The test tube carriers will slide up into the horizontal position (horizontal rotor only) and
the unit will accelerate to the current set speed.
11. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there
are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button
immediately and refer to the section on troubleshooting.
(continued on next page)
page 6