WONDFO Finecare FIA Meter Plus FS-113 Operation Manual Download Page 14

The 

Finecare™ FIA Meter Plus

 has a built-in quality control which aims to monitor the internal system 

performance of the meter. . 
The quality control should be run on and initial use and each day of patient testing. 
To run the Quality Control:
1) Navigate to the main interface, click “QC”. The test device tray will protrude. If the tray does not 
    protrude automatically, click “out”.
2) Gently insert the quality control Card into the test device tray, push the tray inside the meter until 
    resistance is felt. 
3) Click “Start”.

4) Wait until the blue progress bar runs to the end. The result of Quality Control will be displayed as below.
    If the result of quality control is not shown as below, contact the distributor or Guangzhou Wondfo and 
    report the problem 

5) Click “Reset the slot”,to draw back the test device tray. 
6) Click “Print” to make a printed copy of the quality control result if needed. 

Maintenance

Besides regular cleaning or replacing print paper roll when needed, 

Finecare™ FIA Meter Plus

 needs 

minimal maintenance. The instrument shall be plugged out prior to cleaning.

Cleaning

Use a damp cloth with 0.5% bleach, 70% isopropyl alcohol or 70% ethanol to clean the external surface of 

the instrument. Strong bleach (bleach concentration over 0.5%), oxidizing substances or solvents can cause 

permanent damage to the screen and other external surface material thus are prohibited to be applied to the 
meter.

If the meter has been used in clinical setting and needs to be repackaged, replaced or transported, it must be 

cleaned.  Use disinfectant (e.g. bleach concentration lower than 0.1%) and a cloth to wipe external surface 

of the instrument thoroughly. Spray-washing or clean any internal parts and inner surface with disinfectant 
are prohibited as liquid may seep inside the meter.
 

Return

Should a malfunction occur, please contact Guangzhou Wondfo Biotech Co., Ltd. or the local distributor. 
If it were determined that the meter should be returned, a return authorization number will be assigned and 
a replacement meter will be sent to you by Guangzhou Wondfo Biotech Co., Ltd.. The return authorization 
number is printed on the packaging of the replacement meter. The user is expected to utilize the packaging 
of the replacement meter to return the malfunctioning meter. Please send the meter back to Guangzhou 
Wondfo Biotech Co., Ltd. as soon as possible following the receipt of the replacement meter.

Transportation and storage

The original shipping container should be used to transport 

Finecare™ FIA Meter Plus

 or to store it when 

it has not been in use over an extended period. Keep the 

Finecare™ FIA Meter Plus

, dry in upright 

position and in avoidance from sun, mechanical shocks, during transportation and storage. 

Disposal

Any used test device should be disposed according to local ordinances regarding the disposal of 
bio-hazardous materials. 
The 

Finecare™ FIA Meter Plus, 

accessories and the packaging have to be disposed of waste correctly at 

the end of usage. Please follow Local Ordinances or Regulations for disposal.

Page 

24

Page 

25

Section VIII Quality Control

Section IX  Maintenance and Care

Summary of Contents for Finecare FIA Meter Plus FS-113

Page 1: ...Version 30 06 2016 Finecare FIA Meter Plus Operation Manual Model No FS 113...

Page 2: ...Page 3 4 7 8 9 13 17 24 25 26 27 Content Introduction Contents Specifications Warning Precautions and Limitations Installation Operation Communication Settings Quality Control Maintenance and Care Tro...

Page 3: ...more energy the fluorescent dye gives off the stronger the signal After a buffer mixed sample is added to the test device the test device is inserted into the Finecare FIA Meter Plus The meter measure...

Page 4: ...Section Contents 1 List of Contents No 1 2 3 4 5 6 7 8 9 Quantity 1 1 1 1 1 1 1 1 1 Item Finecare FIA Meter Plus Power adapter including the power cord Quality Control Card Ethernet cable Printer pape...

Page 5: ...ity Size Weight Miscellaneous Temperature Humidity Atmospheric pressure Location Temperature Humidity 10 50 14 122 85 10 30 50 86 10 80 86kPa 106kPa Dry clean flat horizontal surface away from direct...

Page 6: ...ou Wondfo Biotech Co Ltd or its representative Do not place objects on Finecare FIA Meter Plus This may damage the optical alignment in turn result in compromised performance or mechanical damage The...

Page 7: ...box 4 2 Set Up Temperature The symbol and the temperature shown below is an indicator of the temperature inside the meter 1 Click set up on the main interface 2 Input q1 in the Authorization codes fi...

Page 8: ...r remain inserted 2 Navigate to the main interface 3 Click Read ID Chip A message said Warning The ID chip ready will prompt click Yes It indicates that the ID chip can be recognized by the meter 4 Ch...

Page 9: ...r the tested item the test result the sample type Name of the patient and the time when the test is conducted are shown for each record If records are more than the screen can show slide up or down on...

Page 10: ...LIS or a PC immediatedly after the test completes 2 history data stored in the meter to be sent to the LIS or a PC Transfer Protocol Instruction Finecare FIA Meter Plus sends data to outside by UDP cl...

Page 11: ...access the Internet one end of the network cable connects Finecare FIA Meter Plus s Ethernet port the other end connects PC s Ethernet port of new Installed NIC Alter the new NIC IP address in compute...

Page 12: ...ollowing information in the TCP IP address window Click Add See Fig 6 9 IP address 192 168 0 88 Subnet mask 255 255 255 0 4 Return Advanced TCP IP settings add gateway by clicking Add as indicated in...

Page 13: ...he computer terminal See Fig 6 17 3 Open LIS data receiving software installed in the computer select network port Click Start See Fig 6 18 Note Choose UDP as communication networks Don t choose Seria...

Page 14: ...been used in clinical setting and needs to be repackaged replaced or transported it must be cleaned Use disinfectant e g bleach concentration lower than 0 1 and a cloth to wipe external surface of th...

Page 15: ...r ID chip Computational Abnormality Mechanical Failure Buzzing Noise Software Abnormality Electrostatic Influence Electric Circuit Failure Test device or ID chip was not inserted The Lot Number on the...

Page 16: ......

Reviews: