The
Finecare™ FIA Meter Plus
has a built-in quality control which aims to monitor the internal system
performance of the meter. .
The quality control should be run on and initial use and each day of patient testing.
To run the Quality Control:
1) Navigate to the main interface, click “QC”. The test device tray will protrude. If the tray does not
protrude automatically, click “out”.
2) Gently insert the quality control Card into the test device tray, push the tray inside the meter until
resistance is felt.
3) Click “Start”.
4) Wait until the blue progress bar runs to the end. The result of Quality Control will be displayed as below.
If the result of quality control is not shown as below, contact the distributor or Guangzhou Wondfo and
report the problem
5) Click “Reset the slot”,to draw back the test device tray.
6) Click “Print” to make a printed copy of the quality control result if needed.
Maintenance
Besides regular cleaning or replacing print paper roll when needed,
Finecare™ FIA Meter Plus
needs
minimal maintenance. The instrument shall be plugged out prior to cleaning.
Cleaning
Use a damp cloth with 0.5% bleach, 70% isopropyl alcohol or 70% ethanol to clean the external surface of
the instrument. Strong bleach (bleach concentration over 0.5%), oxidizing substances or solvents can cause
permanent damage to the screen and other external surface material thus are prohibited to be applied to the
meter.
If the meter has been used in clinical setting and needs to be repackaged, replaced or transported, it must be
cleaned. Use disinfectant (e.g. bleach concentration lower than 0.1%) and a cloth to wipe external surface
of the instrument thoroughly. Spray-washing or clean any internal parts and inner surface with disinfectant
are prohibited as liquid may seep inside the meter.
Return
Should a malfunction occur, please contact Guangzhou Wondfo Biotech Co., Ltd. or the local distributor.
If it were determined that the meter should be returned, a return authorization number will be assigned and
a replacement meter will be sent to you by Guangzhou Wondfo Biotech Co., Ltd.. The return authorization
number is printed on the packaging of the replacement meter. The user is expected to utilize the packaging
of the replacement meter to return the malfunctioning meter. Please send the meter back to Guangzhou
Wondfo Biotech Co., Ltd. as soon as possible following the receipt of the replacement meter.
Transportation and storage
The original shipping container should be used to transport
Finecare™ FIA Meter Plus
or to store it when
it has not been in use over an extended period. Keep the
Finecare™ FIA Meter Plus
, dry in upright
position and in avoidance from sun, mechanical shocks, during transportation and storage.
Disposal
Any used test device should be disposed according to local ordinances regarding the disposal of
bio-hazardous materials.
The
Finecare™ FIA Meter Plus,
accessories and the packaging have to be disposed of waste correctly at
the end of usage. Please follow Local Ordinances or Regulations for disposal.
Page
24
Page
25
Section VIII Quality Control
Section IX Maintenance and Care