Wohler Alarm Profile Editor Installation and Operation
Section 2: Operation
(Continued)
Background
The Wohler
Alarm Profile Editor
is a Windows application designed to allow the user of the Wohler Alarm products to be able
to change alarm trip thresholds.
The application targets the following operating systems: WIN 95, WIN 98, WIN 2000, WIN NT4.0 and WIN XP.
The application is distributed with product on CD-ROM, or may be downloaded from the Wohler Web site (see Factory for
details).
A Wintel architecture machine with at least 10M bytes of hard disk space and a spare COM1 port are required for this application
to be installed correctly.
Installation
The installation package is supplied as three files:
• setup.exe
• setup.1st
• wohler_alarm_profile_editor.cab
Load all three files into a temporary directory (c:\temp on most machines) then start the installation process by running setup.exe
(from Windows Explorer, double click on Setup.exe).
The application installer will take you through the process step by step. It is recommended that you accept defaults where
possible. The default installation path is
c:\Program Files\Wohler_Alarm_Profile_Editor\
.
Application Operation
Execute the
Alarm Profile Editor
by selecting :
Start Menu : Programs : Wohler_Alarm_Profile_Editor :
Wohler_Alarm_Profile_Editor
.
Select
Stereo
or
Mono
mode - depending upon how your unit was ordered.
Note:
For Analog units, Full Scale Input (0 dBFS) is +24dBu.
Example 1:
To set an Analog unit to an Audio Loss threshold of -10 dB ref given a studio reference level of 0 dBu, set the
Loss parameter to 34 dB in the
Alarm Profile Editor
.
Example 2:
To set an Analog unit to an Audio Loss threshold of -20 dB ref given a studio reference level of +4 dBu, set the
Loss parameter to 40 dB in the
Alarm Profile Editor
.
Stereo: Alarm Profile Editor Screen
In
Stereo
, the
Main Editor Screen
is similar to the following: