2.1 Chart Light Does Not Come On
Chapter 2: Troubleshooting Procedures
2.1 Chart Light Does Not Come On
SHOCK HAZARD:
Use care when checking live voltages. Do not touch live terminals,
wires, and ground. Failure to use caution could result in personal injury.
1. All other Integris® 2001 Headwall System electrical systems are
operational.
Yes
No
↓
→
Restore electrical service from the building to the Integris®
2001 Headwall System.
2. If so equipped, the applicable Integris® Headwall 2001 Headwall System
circuit breaker is in the ON position.
Yes
No
↓
→
Reset the applicable circuit breaker to the ON position.
3. The chart light comes on.
Yes
No
↓
→
4. Go to “Final Actions” on page 2-5.
5. Perform the following:
a. Gain access to the chart light.
b. Remove the chart light bulb.
A voltage shows that 109 to 125V AC is available between the bulb center
terminal and ground.
Yes
No
↓
→
6. Replace the chart light bulb (refer to procedure 4.6).
This solves the problem.
Yes
No
↓
→
7. Go to “Final Actions” on page 2-5.
8. Perform the following:
a. Gain access to the chart light switch.
Page 2 - 6
Integris® 2001 Headwall System Service Manual (man148ra)