Safety and performance
Declaration: electromagnetic emissions and immunity for equipment and systems that are not life-supporting and are specified
for use only in a shielded location.
Declaration – electromagnetic immunity
Withings Body Comp is intended for use in the electromagnetic environment specified below.
The customer or the user of the Withings Body Comp should ensure that it is used in such an environment.
Immunity test
IEC 60601 Test Level
Compliance Level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150kHz to 80MHz
N/A
N/A
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
N/A
Portable and mobile RF communications equipment should be used
no closer to any part of the equipment or system including cables than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter. Interference may occur
in the vicinity of equipment marked with the following symbol: .
Electrostatic discharge (ESD)
IEC 61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Floors should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material the relative humidity should be at least 30%Floors
should be wood, concrete, or ceramic tile. If floors are covered with synthetic
material the relative humidity should be at least 30%.
Power frequency (50/60Hz)
magnetic field IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
The power frequency magnetic field should be at levels characteristic of
a typical location in a typical commercial or hospital environment.