WINTACT WT100A Instruction Manual Download Page 5

05

06

5.1.

 

Battery installation;

   

First insert the probe into host probe socket, and install 

batteries in battery compartment, pay attention to the polar of 

batteries. Check battery power after powering on; if battery 

power is low, please charge them in time.

     

5.2.

 

Power on/off;

   

Press "       " to turn on the instrument and enter into sound 

speed measurement interface. At this point, the measurement 

can be started. (As shown), long press "       " button to manually 

shut down.

Measurement interface 
    of sound velocity 

5

Operation preparation

6.1. Apply coupling agent evenly to the measured points on the 

       surface of tested object.

6.2. Thickness can be measured by gently pressing the probe 

       on the object surface with coupling agent.

6

Measurement

MENU

MENU

Note: Please do not install non-rechargeable battery.

7.2. Introduction to main menu functions;
1

Measurement options;

  

Make measurement settings under this menu, including:

    a. Sound speed of measurement. (sound speed bar of 
        measurement interface)  
    b. Measurement unit of   
    c. Measurement resolution

2

Probe

  

Under this menu, select the probe to use in measurement 

   (selected probe icon in measurement interface).

3

Limit value

  

Make limit value settings under this menu.

    a. Set high and low limit value of measurement  
    b. Turn on/off sound alarm for exceeding limit values.

4

Storage

  

Storage related operations can be performed under this menu:

    a. Select the file to save when saving data (the selected 
        file label of the measurement interface)  
    b. View saved data by file  
    c. Delete the data of the selected file

7.1.

 

Introduction to main menu operation;

  

a

.

Enter into main menu

    

Press "       " button in measurement 

       interface to enter into main menu.

  

b

.

Select menu item

    

Press UP or DOWN button under 

       menu interface to select menu item.

  

c

.

Enter into menu item

    

Press OK to enter into selected 

       menu item

  

d

.

Exit menu item

    

Press ESC to return to the previous menu

7

Main menu operation and function introduction

Main menu

MENU

Summary of Contents for WT100A

Page 1: ...eserves the right to modify product design and specification without notice Version WT100A WT130A EN 02 MODEL WT100A WT130A Ultrasonic Thickness Gauge MENU OK VEL UP ESC DOWN Ultrasonic Thickness Gaug...

Page 2: ...nctions 06 Measurement options 06 Probe 06 Limit value 06 Storage 06 Chart 07 Settings 07 12 10 Technology of measurement Menu interface display instruction 03 8 07 Bottom menu operation and function...

Page 3: ...tion and other testing areas Any material that allows ultrasonic to transmit at a constant speed and that is distinguishable from waves reflected from the back side is applicable to this instrument 1...

Page 4: ...be Measurement parameters 1 0 300 0mm 1 0 50 0mm 1 2 300 0mm 1 2 300 0mm 1 0 225 0mm WT130A WT100A Measurement parameters Measuring range Data storage Probe selection WT130A 1 00 300 0mm 1500 WT100A 1...

Page 5: ...ment settings under this menu including a Sound speed of measurement sound speed bar of measurement interface b Measurement unit of c Measurement resolution 2 Probe Under this menu select the probe to...

Page 6: ...to bottom menu functions a Storage Save current measured values to the selected file b Delete Delete measured values and measurement information except for limit values c Calibration Refer to calibra...

Page 7: ...be as smooth as possible through sanding grinding polishing rasping or using high viscosity coupling agent 12 3 Surface made by crude machinery Regular fine grooves on surfaces made by crude machiner...

Page 8: ...h a steel sample with thickness of 4 0 mm Please refer to calibration instruction in item 9 of this manual for calibration method 13 1 Ultra thin material With any ultrasonic thickness gauge if the th...

Page 9: ...omly mixed coupling agent light machine oil etc are suitable For rough surface vertical surface and top surface or aluminum higher viscosity coupling agents such as glycerin creams greases etc can be...

Page 10: ...sides the maintenance fee will be charged according to our company s regulations The maintenance fee for special configuration out of our company s regular products will be charged according to releva...

Reviews: