background image

 

 

WINNCARE   

 

- 5 - 

MAINTENANCE/DISINFECTION 

 

 

It is essential to disconnect the power cord connected to the compressor from the mains 

power outlet. The “mains” indicator must be off.

 

- Do not use a high pressure jet to clean an AXTAIR line compressor 

- Do not spray liquid to avoid any damage to the compressor by penetration of liquid 
inside the box.   

 Do not use abrasive cleaning products such as industrial detergents, acetone type 

solvents,  ether  and  colouring  products  (iodized  alcohol,  potassium  permanganate, 

silver nitrate etc.). Do not allow abrasive materials such as steel wool or “

scotch brite

” 

pads. 

COMPRESSORS: 

Use  a  cloth  soaked  in  a  detergent  product  solution  or  CE*  labelled  surface 
detergent/disinfectant at the use concentrations recommended by the manufacturer. 
Adhere to the remanence times. 

MATTRESS: 

The  pneumatic  unit,  cells  and  harness,  and  the  AXENSOR®  box  for  mattresses  in  the 
AXENSOR® range can easily be dismantled from the base. 
 

For  AXENSOR®  mattresses,  the  AXENSOR®  pneumatic  unit  and  control  box

  must  be 

manually disinfected: use a cloth soaked in a detergent solution or CE* labelled surface 
detergent/disinfectant at the use concentrations recommended by the manufacturer. 
Adhere to the remanence times. 
 

The covers and the bases 

can be machine washed in the following conditions: 

 

Wash in water, maximum temperature 90°C, reduced mechanical 
action, rinsing at decreasing temperature, reduced spinning. 

  Bleaching possible, chlorination at 5000 ppm allowed. 

  Ironing excluded. 

 

Dry  cleaning  not  allowed,  use  of  solvent-based  stain  remover  not 
allowed. 

  Drum drying authorized, moderate temperatures 

 

 

 

 

 

Summary of Contents for Axtair Automorpho Plus

Page 1: ...WINNCARE 1 Technical manual PA0010143 Updated on 10 11 20...

Page 2: ...rence VAXT6 POMPE AUTO 14 Axtair XXL compressor Reference VAXT POMPE XXL 16 Axtair ONE PLUS mattress Reference VAXT MA ONE 18 Axtair AUTO PLUS mattress Reference VAXT MA AUTOP 20 Axtair AUTO PLUS matt...

Page 3: ...the need for servicing Alarm codes Description Meaning Lock indicator steady System problem Support Lock indicator flashing Compressor system problem Dynamic indicator flashing No pressure Treatment...

Page 4: ...eck that the mattress is properly connected to the compressor 2 Check that the CPR valve is well sealed 3 Check the leak tightness of the mattress The compressor measures a nil pressure when pumping 1...

Page 5: ...ectant at the use concentrations recommended by the manufacturer Adhere to the remanence times MATTRESS The pneumatic unit cells and harness and the AXENSOR box for mattresses in the AXENSOR range can...

Page 6: ...EN 14476 A1 NO NO YES YES Unpacking the device Cover machine washing at 60 C with detergent Other surface maintenance bio cleaning using Detergent Disinfectant Replacement cover identical to the orig...

Page 7: ...pliant with the leak tightness test if the difference between P0 and P1 P0 P1 is less than or equal to 5 mBar In the event of non conformity it is necessary to identify the failing components cells co...

Page 8: ...LED Means product servicing is required 17500 hours of operation Servicing compressors at least every 2 years of use is recommended To facilitate the management and scheduling of this servicing counte...

Page 9: ...August 2005 AFTER 13 AUGUST 2005 FOR ELECTRIC AND ELECTRONIC EQUIPMENT PUT ON THE MARKET Winncare France complies with applicable European regulations Winncare France undertakes to finance and organi...

Page 10: ...WINNCARE 10 SPARE PARTS AXTAIR ONE PLUS COMPRESSOR REFERENCE VAXT POMPE ONEP...

Page 11: ...WINNCARE 11 List of references VAXT POMPE ONEP...

Page 12: ...WINNCARE 12 AXTAIR AUTO PLUS COMPRESSOR REFERENCE VAXT POMPE AUTOP...

Page 13: ...WINNCARE 13 List of references VAXT POMPE AUTOP...

Page 14: ...WINNCARE 14 AXTAIR AUTOMORPHO AXENSOR COMPRESSOR REFERENCE VAXT6 POMPE AUTO...

Page 15: ...WINNCARE 15 List of references VAXT6 POMPE AUTO...

Page 16: ...WINNCARE 16 AXTAIR XXL COMPRESSOR REFERENCE VAXT POMPE XXL...

Page 17: ...WINNCARE 17 List of references VAXT POMPE XXL...

Page 18: ...WINNCARE 18 AXTAIR ONE PLUS MATTRESS REFERENCE VAXT MA ONE...

Page 19: ...WINNCARE 19 List of references VAXT MA ONE...

Page 20: ...WINNCARE 20 AXTAIR AUTO PLUS MATTRESS REFERENCE VAXT MA AUTOP...

Page 21: ...WINNCARE 21 List of references VAXT MA AUTOP...

Page 22: ...WINNCARE 22 AXTAIR AUTO PLUS MATTRESS REFERENCE VAXT MA AUTOP100...

Page 23: ...WINNCARE 23 List of references VAXT MA AUTOP100...

Page 24: ...WINNCARE 24 AXTAIR AUTO PLUS MATTRESS REFERENCE VAXT MA AUTOP120...

Page 25: ...WINNCARE 25 List of references VAXT MA AUTOP120...

Page 26: ...WINNCARE 26 AXTAIR AUTOMORPHO AXENSOR AT12 MATTRESS REFERENCE VAXT6 MA AUTO...

Page 27: ...WINNCARE 27 List of references VAXT6 MA AUTO...

Page 28: ...WINNCARE 28 AXTAIR AUTOMORPHO AXENSOR AT15 MATTRESS REFERENCE VAXT6 MA MAX...

Page 29: ...WINNCARE 29 List of references VAXT6 MA MAX...

Page 30: ...WINNCARE 30 AXTAIR AUTOMORPHO AXENSOR AT20 MATTRESS REFERENCE VAXT6 MA AT20 90...

Page 31: ...WINNCARE 31 List of references VAXT6 MA AT20 90...

Page 32: ...WINNCARE 32 AXTAIR AUTOMORPHO AXENSOR AT20 MATTRESS REFERENCE VAXT6 MA AT20 80...

Page 33: ...WINNCARE 33 List of references VAXT6 MA AT20 80...

Page 34: ...WINNCARE 34 AXTAIR XXL MATTRESS REFERENCE VAXT MA 120XXL...

Page 35: ...WINNCARE 35 List of references VAXT MA 120XXL...

Page 36: ...WINNCARE 36 AXTAIR XXL MATTRESS REFERENCE VAXT MA 140XXL...

Page 37: ...WINNCARE 37 List of references VAXT MA 140XXL...

Page 38: ...Phone 33 0 4 66 02 15 15 Fax 33 0 4 66 02 15 00 Email contact askle com www winncare fr...

Reviews: