Guidance and manufacturer’s declaration-electromagnetic emissions
The W32L100-PTA1 isintended for useinthe electromagnetic environmentspecified below.
Thecustomeror theuseroftheW32L100-PTA1 should assure that itisused insuchanenvironment.
Emissiontest
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group1
The W32L100-PTA1 uses RF energy only for its internal
function.Therefore,its RF emissions areverylowand arenot
likelytocause anyinterference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The W32L100-PTA1 is suitableforuseinall establishments,
including domestic establishmentsand those directlyconnected
tothe public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Guidanceandmanufacturer’sdeclaration -electromagnetic immunity
The W32L100-PTA1 isintended for use in theelectromagnetic environment specifiedbelow.The customerorthe user ofthe
W32L100-PTA1 should assure that it isusedinsuchasenvironment.
Immunitytest
IEC60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshould be wood, concreteor
ceramic tile. Iffloorsare coveredwith
synthetic material, the relative
humidity shouldbe at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kVfor power supplylines
±1 kVfor input/ output lines
±2kVfor power supplylines
±1kVfor input/ output lines
Mains power qualityshould be that ofa
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kVline(s) to line(s)
±2 kVline(s) to earth
±1kVdifferential mode
±2kVcommon mode
Mains power qualityshould be that ofa
typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage
variationsonpowersu
pply input lines
IEC 61000-4-11
<5%
(>95% dip in) for0.5cycle
40%
(60% dipin) for5cycles
70%
(30% dipin) for 25 cycles
<5%
(>95% dip in) for5sec
<5%
(>95% dip in) for0.5cycle
40%
(60% dipin) for 5 cycles
70%
(30% dipin) for 25 cycles
<5%
(>95% dip in) for5sec
Mains power qualityshould be that
ofa typical commercial or hospital
environment. Iftheuser ofthe-
W32L100-PTA1 requirescontinued
operation during power mains
interruptions, it isrecommended that
the W32L100-PTA1be poweredfrom
anuninterruptiblepowersupplyora
battery.
Power frequency
(50/60Hz)
magneticfield
IEC 61000-4-8
3 A/m
3A/m
Power frequencymagnetic
fieldsshould be at levelscharacteristic
ofa typical locationin a typical
commercial or hospitalenvironment.
NOTE: istheA.C.mainsvoltageprior toapplicationofthetestlevel.