- 26 -
Guidance and
manufacturer’s declaration-electromagnetic emissions
The M270TF-XXX / M320TF-XXX is intended for use in the electromagnetic environment specified below.
The customer or the user of the M270TF-XXX / M320TF-XXX should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The M270TF-XXX / M320TF-XXX uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The M270TF-XXX / M320TF-XXX is suitable for
use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and
manufacturer’s declaration - electromagnetic immunity
The M270TF-XXX / M320TF-XXX is intended for use in the electromagnetic environment specified below.
The customer or the user of the M270TF-XXX / M320TF-XXX should assure that it is used in such as
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/ output
lines
±2 kV for power supply
lines
±1 kV for input/ output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5%
(> 95% dip in) for 0.5 cycle
40%
(60% dip in) for 5 cycles
70%
(30% dip in) for 25 cycles
< 5%
(> 95% dip in) for 5 sec
< 5%
(> 95% dip in) for 0.5 cycle
40%
(60% dip in) for 5 cycles
70%
(30% dip in) for 25 cycles
< 5%
(> 95% dip in) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the LMD- M270TF-
XXX / M320TF-XXX requires
continued operation during
power mains interruptions, it
is recommended that the
M270TF-XXX / M320TF-
XXXbe powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: is the A.C. mains voltage prior to application of the test level.