5. NCV measurement;
The operation is as follows:
5-1. Press and hold the

NCV/LIVE

button; enter the EF measurement.

5-2. The front end of the meter has NCV test points. As long as the point is close to the AC voltage, the buzzer will emit different continuations

according to the different strength of the signal, and the LCD will also display according to the strength of the signal. Different number of segments.

5-3.Cycle to trigger "NCV/LIVE" key, enter live live live live wire measurement, display live character on LCD, close to live wire test point with a

red meter, and make reliable contact with the point, the black meter end is suspended, and does not contact any test point; if the test point is live wire, the
LCD will display OL, and the buzzer will send continuous sound.

6.Frequency measuremen

Input sensitivity: 1V rms; overload protection: 550V DC or AC peak (no more than 10 seconds)
Frequency measurement
1.trigger power key; switch to frequency measurement function;
2.The black test lead is negative, the red test lead is positive, and the test lead reliably touches the test point;
3 Measurement values will be displayed on the LCD.
Note:
1) When the input exceeds 10Vrms, it can be read, but the error may be larger;

2) In a noisy environment, it is best to use a shielded cable when measuring small signals;

Accuracy

Range

6000counts

Resolution

10Hz

±(

0.1%+3

0.01Hz

100Hz

0.1Hz

1kHz

1Hz

10kHZ

10Hz

100kHz

100Hz

1MHz

1kHz

10MHz

10kHz

Summary of Contents for ET8138

Page 1: ...of Pocket Type Intelligent Scanning Digital Multimeter Index 1 Overview 1 2 Safety precautions 1 3 Characteristics 1 4 Operation panel instructions 3 5 Technical Specifications 4 6 Automatic switch ma...

Page 2: ...superior tool and an ideal tool for laboratories factories radio enthusiasts and families 2 Safety Precautions This series of instruments is designed in accordance with IEC1010 Safety Standards promul...

Page 3: ...haracteristics 2 1 Accuracy a of reading least significant digit to ensure accuracy Ambient temperature 23 5 relative humidity 75 calibration guarantee period from the date of manufacture for one year...

Page 4: ...frequency and buzzer Black is the negative end of the input Note 1 The back light is always on LCD is highlighted and the user is constantly on In use the instrument will not turn off the power supply...

Page 5: ...the red test lead is greater than 0 8V regardless of the AC voltage or the DC voltage the meter will compare the DC component and the AC component take the larger component signal and then according...

Page 6: ...if the voltage in the loop is greater than 0 8V the meter will mistake it for voltage measurement and enter the electricity measurement mode 2 5 When measuring the value at the negative end of the bl...

Page 7: ...ill be displayed on the LCD The capacitance measurement range is 10nF 100nF 1uF 10uF 100uF 1mF 10mF 60mF Note 1 When measuring capacitance with 10nF file there may be residual reading on the screen di...

Page 8: ...e contact with the point the black meter end is suspended and does not contact any test point if the test point is live wire the LCD will display OL and the buzzer will send continuous sound 6 Frequen...

Page 9: ...completely welded at the input end there will be certain data on the LCD In the design we define the manual DC voltage level the input end has 10 words the AC voltage level has 100 words and the capac...

Reviews: