Setting Step
Size and
Dwell Time
There are two ways to set the size of each step of
the step sweep—set the step size or set the number
of steps. The step size range is 1 kHz to the full fre-
quency range of the sweep generator (0.1 Hz to full
frequency range with Option 11); the number of
steps range is 1 to 10,000. The dwell time-per-step
of the step sweep can be set for any time in the
range of 1 ms to 99 sec. The step size and dwell
time-per-step parameters are set from the step
sweep ramp menu.
To go to the Step Sweep Ramp menu (shown below)
from the Step Sweep menu, press
Sweep Ramp
.
This menu lets you set the dwell time, the step size,
the number of steps, and go to the trigger menu.
Press
Dwell Time
to open the dwell time-per-step
parameter.
Press
Step Size
to open the step size parameter.
Press
Num of Steps
to open the number of steps pa-
rameter.
Open the parameter you wish to change, then edit
the current value using the cursor control key or the
rotary data knob or enter a new value using the key
pad and appropriate termination key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or make another menu
selection.
To go to the Step Sweep Trigger menu from this
menu, press
Trigger Menu
. The trigger menu lets
you select a sweep trigger (previously described on
page 3-24).
Press
Previous Menu
to return to the Step Sweep
menu.
LOCAL (FRONT
SWEEP FREQUENCY
PANEL) OPERATION
OPERATION
RANGE ERR
This error message is displayed
when (1) the step size value entered
is greater than the sweep range or
(2) the number of steps entered re-
sults in a step size of less than 1 kHz
(0.1 Hz with Option 11). Entering a
valid step size will clear the error.
3-26
681XXA OM