meters
ID
range
setting
Alarm too high
temperature
P1
0–99.9
℃
38.8
℃
Heating stop
P2
0–99.9
℃
37.8
℃
Please note that the radiator will
switch off when reaching the
entered value, but will still radiate
some warmth, and the temperature
might still rise a little bit in the incub-
ator. E. g. this means that a temper-
ature of 38
can be reached,
℃
when setting the temperature to
37.8
. If you wish to avoid this, the
℃
temperature of the heating stop
needs to be corrected downward.
Alarm too low
temperature
P3
0–99.9
℃
36
℃
Table 1: Adjustment of the finer parameters via the “set” and “+” button
3. “Set” and “–”: Holding both at the same will open the menu for adjusting the turning
options.
The display will show the individual parameter, displayed as number/letter combina-
tion (parameter ID).
With the “+” and “–” buttons, you can choose the various parameters.
If you wish to alter a parameter, push the “set” button, then alter the value up or down-
ward via the “+” and “–” buttons.
Pushing the “set” button again, will save the changes.
Turning para-
meters
Parameter
ID
Adjustment range
Standard
setting
Note
Turning fre-
quency
F1
0–999 min
90 min
When adjusting to 0, no
turning takes place.
Turning dura-
tion
F2
0–999 sec
10 sec
When adjusting to 0, no
turning takes place.
Table 2: Adjusting the turning via the “set” and “–” buttons
4. “+” and “-”: Holding both pushed at the same time for a longer duration (approx. 8 sec)
resets the incubator back to factory settings, there will be a peep sound