radiate some warmth, and the
temperature might still rise a
little bit inside the incubator.
This means e. g. that a temper-
ature of 38°C can be reached
when setting the temperature
to 37.8°C. If you wish to avoid
this, the heating stop temperat-
ure needs to be corrected
downwards.
Heating start
P3
0–99.9°C
37.5°C
Temperature alarm
too low
P4
0–99.9°C
37°C
Humidity alarm too
high
H1
0–99%
80%
Humidity alarm too
low
H2
0–99%
40%
Table 4: Adjustment via “set” and “+” button for finer parameters
“+” : Allows for the turning to be carried out manually.
Push the buttons once shortly: The device will turn the eggs for the time set in dis-
play 1. After that, the counter will reset itself to the adjusted time (in this case 2 hours)
and will turn the eggs again after 2 hours.
The turning frequency can be adjusted, see table 3 further up.
“+” and “–”: Holding both down at the same for a longer moment (approx. 8 seconds)
will reset the incubator to factory settings, a beep sound will sound.
Display, function buttons and basic settings
For the incubator with item no. 51076
On the left outer corner of the control panel is the connection for the electricity plug.
On the right outer corner there is the on/off switch. It is important that the electricity plug is completely
plugged in before operating the on/off switch. If the on/off switch is glowing, the incubator is switched
on and ready for use.
Besides the four displays (1 and 2), there are three more buttons with which the incubator is operated
(see
“Function buttons description”)
and two small lamps which glow when the heater is heating
(“work”) or adjustments are being made (“set”).
Display 1 shows the temperature in Celsius.
Display 2 shows the humidity percentage.
© by WilTec Wildanger Technik GmbH Item 50039, 50041, 51074–51077, 51271–51274
Page 13
http://www.WilTec.de
http://www.aoyue.eu
11 2020
-1