WILDLIFE 265A Quick Start Manual Download Page 3

IOP 006A.1 

Effective Date: 1/09/2017 

Page 3 of 4 

 

III. 

StatFace ™ electrode monitoring 

 

A. 

The StatFace symbols in the upper left corner of the display indicate the state of the 

electrode: 

 

 

 

1. 

  

Electrode is calibrated and in good working order. 

 

 

2. 

  

Electrode response is slow/deteriorating.  Electrode should be  

 

 

cleaned and recalibrated. 

 

 

3. 

  

Electrode response is very slow and correct measurement is no  

 

 

longer ensured.  Clean or replace electrode 

 

IV. 

Data logging mode 

 

A. 

Press the 

STO

 key and then the clock key to access data log mode options.  Press the up 

or down arrow key to move between options.  Pressing 

STO

 also advances to the next 

menu item.  Pressing the 

meas

 key at any time ends data logging mode. 

 
B. 

To set parameters scroll to 

PAr

: Press 

STO

 and Int (time interval between logged 

 

data points) is displayed 

 

1. 

In 

INT

, press 

STO

 to access interval range, which should be set at 60:00 (60 

minutes); if not, adjust interval using up or down arrows.  Press 

STO 

again and 

display returns to 

Cont

 
C. 

Scroll to 

Strt

 (start):  Pressing 

STO

 clears the logger memory and begins data logging. 

The first logged data point is in memory location “00”. 

 

D. 

The Orion model 265A and 266S records up to 100 data points (memory locations “00” 

to “99”).  The data log does not automatically stop after reaching 100 data points.  It 

continues logging beginning at location “00” and overwrites previously recorded data. 

 

V. 

Deployment 

 

A. 

Place the unit in an inconspicuous location on the stream bank so the probe can be placed 

in moving water several feet from the bank. 

 

B. 

Suspend the probe in the water column (e.g. drape from a tree branch, place on a forked 

stick stuck into the stream bed or on a stack of rocks).  This minimizes interference from 

silt or biological accumulation on the electrode.   

 
C. 

Enter data logging mode (press 

STO

 then 

clock

), move to 

Strt

, then press 

STO

 to 

begin logging.  Record the time logging started, logger pH and water temperature on the 

pH Logger Calibration and Deployment Record form.   

 

VI. 

Retrieval  

 
B. 

Press 

meas

 to end data logging and turn the unit off. 

 

C. 

Downloading pH logger data with Paraly SW109 data transfer software 

 

1. 

Connect the logger to a computer with the serial interface cable.  The 3-prong 

end is connected to the logger with the “PC” labeled side facing the front of the 

Summary of Contents for 265A

Page 1: ...Oceans Sea Lamprey Control Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada and U S Fish and Wildlife Service Lake Champlain Fish and Wildlife Resources Office 11 Lincoln St Essex Junction VT 05452 INSTRUMENT OPERATING PROCEDURE INSTRUMENT Portable pH measurement and logging device MODEL 265A 266S MANUFACTURER Thermo Orion PRECAUTIONS POTENTIAL INTERFERENCES Assure that the pro...

Page 2: ...se probe in pH 7 buffer stir buffer for several seconds to remove bubbles from electrode and wait a minimum of 5 minutes for the probe to stabilize Press cal and gently stir to calibrate to pH 7 the measured mV value is displayed When pH 7 calibration is complete CAL 2 is displayed G Rinse probe with distilled water pat off excess water and rinse with the pH 4 buffer Immerse probe in pH 4 buffer s...

Page 3: ...gger memory and begins data logging The first logged data point is in memory location 00 D The Orion model 265A and 266S records up to 100 data points memory locations 00 to 99 The data log does not automatically stop after reaching 100 data points It continues logging beginning at location 00 and overwrites previously recorded data V Deployment A Place the unit in an inconspicuous location on the...

Page 4: ...any salt deposits with distilled water Immerse in electrode storage solution for at least one hour D The electrode can be cleaned by soaking in pH 4 buffer If this does not improve performance use the ph electrode cleaning kit Orion 900020 following instructions in the kit MAINTENANCE All sensors are shipped with the measuring end covered with a wetting cap Sensors are stored in a pH 4 buffer solu...

Reviews: